Medical Devices Regulation (EU) 2017/745

Last reviewed: October 2026 · Sources listed at the end of this page.

Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices — commonly the "MDR" — is the EU framework for the placing on the market and putting into service of medical devices and their accessories. It replaced Directives 93/42/EEC (medical devices) and 90/385/EEC (active implantable medical devices). The MDR was published in the OJEU as OJ L 117, 5.5.2017, p. 1 and has applied from 26 May 2021. The application date was deferred by one year from the original 26 May 2020 by Regulation (EU) 2020/561 in response to COVID-19. Subsequent amendments extended transitional periods for legacy devices certified under the predecessor directives.

At a glance. The MDR is the EU law on CE marking of medical devices and has applied since 26 May 2021. Legacy devices that meet the Article 120 conditions may be placed on the market until 31 December 2027 or 31 December 2028, depending on class. A Notified Body is needed for every class except basic Class I devices (not sterile, no measuring function, not reusable surgical instruments). Next key date: 28 November 2026, for EUDAMED registration of devices already on the market. The Commission’s proposed revision is not yet law.

Legal status and timeline

Scope: products covered

Article 1 and Annex XVI of the MDR apply to:

Software as a medical device

Software intended for medical purposes is a medical device under MDR Article 2(1). The classification rules in Annex VIII apply, including specific Rule 11 which has placed much medical software into Class IIa, IIb, or III depending on the seriousness of the decision the software supports. Rule 11 has been one of the principal sources of difficulty in MDR transition for software developers, as it pushed many products previously self-certified under the MDD into Notified-Body-required classes.

Borderline products

The Commission's Manual on borderline and classification, which is updated periodically, addresses contested cases — software with general wellness purposes vs. medical purposes, accessories vs. consumables, the boundary with cosmetics under Regulation (EC) 1223/2009, with biocides under Regulation (EU) 528/2012, and with personal protective equipment under PPE Regulation 2016/425.

Device classification (Annex VIII)

MDR Annex VIII contains 22 rules dividing devices into four classes by risk:

Essential requirements: general safety and performance

MDR Annex I sets out the general safety and performance requirements (GSPRs), structured into three chapters:

Conformity assessment procedures

Articles 52–53 and Annexes IX, X, XI set the routes:

Notified Body certificates are valid for the period they indicate, which may not exceed five years, and may be extended for further periods of up to five years each following re-assessment (Article 56(2)).

See conformity assessment modules and Notified Bodies. About 50 Notified Bodies were reported as designated under the MDR in 2026; NANDO holds the current list. To check a number on a label, see CE number lookup.

Technical documentation

Annexes II and III set the technical documentation requirements:

The documentation must be retained for 10 years after the last device was placed on the market, or 15 years for implantable devices (Article 10(8)).

EU Declaration of Conformity

Article 19 and Annex IV require the Declaration to include the Basic UDI-DI of the device, the manufacturer's Single Registration Number (SRN), the risk class, references to any common specifications used, and the Notified Body's name, identification number and certificate references. See EU Declaration of Conformity. The Declaration must be translated into the official Union language or languages required by the Member States where the device is made available (Article 19(1)).

Marking and labelling

Article 20 requires the CE marking; the four-digit Notified Body identification number follows the mark wherever a Notified Body took part in the conformity assessment, so basic Class I devices carry no number. Beyond the CE mark, the label must include (Annex I, Section 23):

UDI and EUDAMED

The Unique Device Identification system (Articles 27–30) requires each device to have a UDI consisting of a Device Identifier (DI) and a Production Identifier (PI). The UDI is registered in EUDAMED — the European Database on Medical Devices established under Article 33. EUDAMED has six modules: Actor Registration, UDI/Device Registration, Notified Bodies and Certificates, Clinical Investigations and Performance Studies, Vigilance and Post-Market Surveillance, and Market Surveillance. Under the gradual roll-out introduced by Regulation (EU) 2024/1860, each module becomes mandatory six months after the Commission publishes a notice of its functionality. Commission Decision (EU) 2025/2371 (OJ 27.11.2025) confirmed the Actor registration, UDI/Device registration, Notified Bodies and Certificates, and Market Surveillance modules, which have therefore been mandatory since 28 May 2026; the Vigilance and Clinical Investigations/Performance Studies modules are not yet mandatory, because no notice has been published for them.

Article 123(3) sets the follow-on deadlines, counted from the notice of 27 November 2025:

Person Responsible for Regulatory Compliance (PRRC)

Article 15 requires each manufacturer (and, by Article 15(6), each Authorised Representative) to have at least one person responsible for regulatory compliance with specified qualifications: a diploma in a relevant field (law, medicine, pharmacy, engineering) plus one year's professional experience, or four years' professional experience. Micro and small enterprises are not required to have the PRRC within the organisation but must have a person continuously and permanently at their disposal.

Recent and upcoming changes

Related legislation

Common errors

Frequently asked questions

When do MDD certificates stop being usable?

Devices under a certificate kept valid by Article 120 may be placed on the market until 31 December 2027 (all Class III, and Class IIb implantable devices apart from the listed exceptions) or 31 December 2028 (other Class IIb, Class IIa, Class I sterile or measuring). This holds only if the Article 120(3c) conditions were met, including a formal application to a Notified Body by 26 May 2024 and a signed agreement by 26 September 2024. Units already lawfully placed on the market may continue to be made available.

When must a device already on the market be registered in EUDAMED?

By 28 November 2026, if the device was first placed on the market before 28 May 2026 and units continue to be placed on the market after that date. New devices must be registered before they are placed on the market.

Does a Class I device need a Notified Body?

Not if it is non-sterile, has no measuring function and is not a reusable surgical instrument: the manufacturer draws up the technical documentation and the EU declaration of conformity (Article 52(7)). For Class Is, Im and Ir devices a Notified Body assesses the sterile, measuring or reprocessing aspects only.

Has the MDR revision been adopted?

No. COM(2025) 1023 was proposed on 16 December 2025 and is still in the ordinary legislative procedure. The current MDR, including the five-year limit on certificate validity in Article 56(2), continues to apply.

Are CE-marked devices still accepted in Great Britain?

Yes. On 21 October 2026 the UK Government confirmed long-term acceptance of CE-marked devices, but the legislation had not been made at the time of this review. Northern Ireland continues to apply the EU MDR and IVDR.